Description

MALDI-TOF MS has been shown capable of rapidly and accurately characterizing bacteria. Highly reproducible spectra are required to ensure reliable characterization. Prior work has shown that spectra acquired manually can have higher reproducibility than those acquired automatically. For this reason,

MALDI-TOF MS has been shown capable of rapidly and accurately characterizing bacteria. Highly reproducible spectra are required to ensure reliable characterization. Prior work has shown that spectra acquired manually can have higher reproducibility than those acquired automatically. For this reason, the objective of this study was to optimize automated data acquisition to yield spectra with reproducibility comparable to those acquired manually. Fractional factorial design was used to design experiments for robust optimization of settings, in which values of five parameters (peak selection mass range, signal to noise ratio (S:N), base peak intensity, minimum resolution and number of shots summed) commonly used to facilitate automated data acquisition were varied. Pseudomonas aeruginosa was used as a model bacterium in the designed experiments, and spectra were acquired using an intact cell sample preparation method. Optimum automated data acquisition settings (i.e., those settings yielding the highest reproducibility of replicate mass spectra) were obtained based on statistical analysis of spectra of P. aeruginosa. Finally, spectrum quality and reproducibility obtained from non-optimized and optimized automated data acquisition settings were compared for P. aeruginosa, as well as for two other bacteria, Klebsiella pneumoniae and Serratia marcescens. Results indicated that reproducibility increased from 90% to 97% (p-value [~ over =] 0.002) for P. aeruginosa when more shots were summed and, interestingly, decreased from 95% to 92% (p-value [~ over =] 0.013) with increased threshold minimum resolution. With regard to spectrum quality, highly reproducible spectra were more likely to have high spectrum quality as measured by several quality metrics, except for base peak resolution. Interaction plots suggest that, in cases of low threshold minimum resolution, high reproducibility can be achieved with fewer shots. Optimization yielded more reproducible spectra than non-optimized settings for all three bacteria.

Reuse Permissions
  • Downloads
    pdf (1 MB)

    Details

    Title
    • A Designed Experiments Approach to Optimization of Automated Data Acquisition During Characterization of Bacteria With MALDI-TOF Mass Spectrometry
    Contributors
    Date Created
    2014-03-24
    Resource Type
  • Text
  • Collections this item is in
    Identifier
    • Digital object identifier: 10.1371/journal.pone.0092720
    • Identifier Type
      International standard serial number
      Identifier Value
      1045-3830
    • Identifier Type
      International standard serial number
      Identifier Value
      1939-1560

    Citation and reuse

    Cite this item

    This is a suggested citation. Consult the appropriate style guide for specific citation guidelines.

    Zhang, L., Borror, C. M., & Sandrin, T. R. (2014). A Designed Experiments Approach to Optimization of Automated Data Acquisition during Characterization of Bacteria with MALDI-TOF Mass Spectrometry. PLoS ONE, 9(3). doi:10.1371/journal.pone.0092720

    Machine-readable links